13-2
Warning
For SunTech NIBP module, when monitoring neonates, the monitor cannot
identify the cuff type, so causing excessive cuff pressures on a neonate
when taking the blood pressure measurement for a neonate in the ―Adult‖
patient type setting. It is the responsibility of the user (physician or nurse)
to always determine Adult or Neonate mode when initiating a BP
measurement.
Warning
For KRK NIBP module, even though the monitor can identify the cuff type
so it will stop inflation and indicate
―Cuff error‖ when taking the blood
pressure measurement for a neonate in the
―Adult‖ patient type setting, the
user (physician or nurse) should pay more attention to select the correct
patient type.
Warning
NIBP monitoring is prohibited to those who have severe hemorrhagic
tendency or with sickle cell disease, otherwise, partial bleeding will appear.
Warning
It is recommended to take the blood pressure measurement manually.
Medical staff must be present when performing auto or sequential
measurement.
Warning
DO NOT wrap the cuff on limbs with transfusion tube or intubations or skin
lesion area, otherwise, injury may be caused to the limbs.
Warning
If the patient is moving or suffering tremble, hyperkinesia or arrhythmia, it
may cause the inflation time of inflatable balloon endures longer, which
may not only prolong the measurement time, but also result in the body
wrapped by the cuff is troubled by purpura, hypoxemia and neuralgia
because of the friction.
Warning
The air-hose which connects the cuff and monitor should be straightway
without any tangle.
Caution
Prior to use of the cuff, empty the cuff until there is no residual air inside it
to ensure accurate measurement.
Содержание AIView 12V
Страница 1: ...AIView Series Patient Monitor Operator s Manual...
Страница 2: ......
Страница 6: ...IV This page intentionally left blank...
Страница 24: ...1 7 This page intentionally left blank...
Страница 26: ...2 2...
Страница 31: ...2 7 2 6 2 Rear View 1 2 3 4 1 Handle 2 Device label 3 Equipotential grounding terminal 4 AC power connector...
Страница 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Страница 35: ...2 11 This page intentionally left blank...
Страница 44: ...3 9 1 2 3 4 5 8 9 7 6...
Страница 55: ...3 20 This page intentionally left blank...
Страница 71: ...5 11 This page intentionally left blank...
Страница 78: ...6 7 This page intentionally left blank...
Страница 106: ...8 5 This page intentionally left blank...
Страница 112: ...9 6 This page intentionally left blank...
Страница 119: ...10 7 This page intentionally left blank...
Страница 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Страница 165: ...15 12 Other sources of interference if any...
Страница 166: ...15 13 This page intentionally left blank...
Страница 181: ...17 8 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 185: ...17 12 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 194: ...17 21 This page intentionally left blank...
Страница 204: ...19 5 This page intentionally left blank...
Страница 210: ...20 6 This page intentionally left blank...
Страница 229: ...23 10 This page intentionally left blank...
Страница 248: ...25 8 This page intentionally left blank...
Страница 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Страница 283: ...D 2 This page intentionally left blank...
Страница 289: ...E 6 less than 3V m...