13-4
Patients with severe bleeding tendency or sickle cells are
forbidden to use the NIBP measurement, otherwise local
bleeding may be caused.
Do not apply cuff on the limb where skin damage has occurred
or is expected.
When extremely low or high heart rate or serious arrhythmia of
the patient occurs. Especially auricular fibrillation will lead to
unreliable or impossible measurement.
When the patient is suffering from massive hemorrhage,
hypovolemia, shock and other conditions with rapid blood
pressure change or when the patient has too low body
temperature, the reading will not be reliable, for reduced
peripheral blood flow will lead to reduced arterial pulsation.
For patients with severe coagulation disorders, it is necessary to
determine whether to carry out automatic blood pressure
measurement according to clinical evaluation, because there is
a risk of hematoma at the friction between the limb and the cuff.
13.4 Measurement Modes
The following NIBP measurement modes are supported:
Manual - measurement on demand.
Auto: repeated measurements at the set interval. Measuring at
integral points are available.
STAT: continual rapid series of measurements over a
five-minute period, then return to the previous mode. The
monitor will not stop making measurement until the measuring
time is over 5 minutes or the operator stops it manually.
Sequence: continual automatic measurement at set duration
and interval of each stage.
In this mode, a complete measurement includes up to 7 phases (A,
B, C, D, E, F, G). You can set the measurement duration of each
cycle and the interval between two NIBP measurements.
ABPM (ambulatory blood pressure monitoring): conduct NIBP
measurements according to the set sequential mode.
Warning
―STAT‖ mode can only be used for adults.
Содержание AIView 12V
Страница 1: ...AIView Series Patient Monitor Operator s Manual...
Страница 2: ......
Страница 6: ...IV This page intentionally left blank...
Страница 24: ...1 7 This page intentionally left blank...
Страница 26: ...2 2...
Страница 31: ...2 7 2 6 2 Rear View 1 2 3 4 1 Handle 2 Device label 3 Equipotential grounding terminal 4 AC power connector...
Страница 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Страница 35: ...2 11 This page intentionally left blank...
Страница 44: ...3 9 1 2 3 4 5 8 9 7 6...
Страница 55: ...3 20 This page intentionally left blank...
Страница 71: ...5 11 This page intentionally left blank...
Страница 78: ...6 7 This page intentionally left blank...
Страница 106: ...8 5 This page intentionally left blank...
Страница 112: ...9 6 This page intentionally left blank...
Страница 119: ...10 7 This page intentionally left blank...
Страница 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Страница 165: ...15 12 Other sources of interference if any...
Страница 166: ...15 13 This page intentionally left blank...
Страница 181: ...17 8 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 185: ...17 12 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 194: ...17 21 This page intentionally left blank...
Страница 204: ...19 5 This page intentionally left blank...
Страница 210: ...20 6 This page intentionally left blank...
Страница 229: ...23 10 This page intentionally left blank...
Страница 248: ...25 8 This page intentionally left blank...
Страница 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Страница 283: ...D 2 This page intentionally left blank...
Страница 289: ...E 6 less than 3V m...