23
Leica ST5020
Important!
Prior to taking any further installation
steps, switch the instrument off and
unplug from mains.
• Cover panel (
26
) must be removed from the
right inner wall of the instrument, so that the
transfer arm can access the oven stations.
• To do so, remove six screws (
27
) (Allen key,
size 2.5) and remove cover panel (
26
).
• Store screws and cover panel in safe place
for further use.
• Remove cover panel (
28
) located to the right
of the electrical connections at the rear wall
of the instrument.
• To remove panel (
28
), only remove screws
(
29
).
Do not loosen
the two screws (
30
)!
• Once the oven is installed, attach cover
panel (part of oven kit delivery) (
17
in Fig. 10
)
with four screws (
29
) in such a way that the
voltage selected is visible from the outside.
Fig. 12
Fig. 13
26
27
27
28
29
29
29
29
30
4. Setting up the instrument