54
Instructions for Use, V2.0, RevG - 07/2016
5. Operation
5.5 Instrument status
→
•
INSTRUMENT STATUS
may
also be accessed from the
main menu, in which case
a
BACK
button will be dis-
played in the menu bar.
• The graphical display pro
-
vides information on station
configuration, reagent status
and current location of racks
during a program run.
• Like in
STATION DETAILS
,
programs ready to be start-
ed (with the exception of
programs defined as 'white')
are displayed above the sta-
tion buttons (
1
).
• If one of the colored program buttons (
1
) is pressed, the
PROGRAM-
MING
menu of the corresponding program opens up.
(See
chapter 5.6
for details).
• The
SCHEDULE
button leads to the 'Schedule' menu, where all racks
currently being processed are listed in a table.
(See
chapter 5.7
for details).
• The
PAUSE
button is displayed only while staining programs are in
progress.
Press
PAUSE
to:
interrupt rack processing - a window containing the button
CONTINUE
will open up and alarm type 3 will be triggered. Press
CONTINUE
to re-
sume processing.
Fig. 50
1
INSTRUMENT STATUS is the standard screen of the instrument, displayed automatically when
the instrument is switched on (unless a different language is selected -
s. chapter 4.1
).