What to do if ...?
56
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 57
13.3 Error messages on the control unit
When the control unit detects an error, the yellow "Check" button lights up.
X
Press the "Check" button.
The list with error messages is displayed.
X
To acknowledge a message, select the message and press the "Confirm" button.
When there is no error message pending, the yellow "Check" button disappears.
Message
Cause
Remedy
Luxmeter, no calibration data available
Luxmeter may not be calibrated.
X
Contact your Leica Representative.
Lamp1 overtemperature
The ventilation hole may be clogged/blocked.
X
Remove any blockage from the ventilation hole.
Lamp2 overtemperature
The ventilation hole may be clogged/blocked.
X
Remove any blockage from the ventilation hole.
Lamp1 fan defective
Lamp 1 fan is defective.
X
Contact your Leica Representative.
Lamp2 fan defective
Lamp 2 fan is defective.
X
Contact your Leica Representative.
Lamp door is open
The access door of the illumination unit is not
closed.
The push button for Illumination on/off flashes.
X
Close the access door of the illumination unit and
lock it using the door knob.
Lamp1 defective
Lamp 1 is defective.
X
After the operation replace lamp 1.
Lamp2 defective
Lamp 2 is defective.
X
After the operation replace lamp 2.
Exhaust fan defective
The exhause fan is defective.
X
Contact your Leica Representative.
Overtemperature in tower base
The ventilation hole may be clogged/blocked.
X
Remove any blockage from the ventilation hole.
No lamp selected
The lamp selection switch is in an intermediate
position.
X
Move the knob to a valid position.
Optics carrier not found
• Cable Connection loose.
• Cable damaged.
• Controller defective.
X
Contact your Leica Representative.
CAN-handle left not found
• Cable Connection loose.
• Cable damaged.
• Controller defective.
X
Contact your Leica Representative.
CAN-handle right not found
• Cable Connection loose.
• Cable damaged.
• Controller defective.
X
Contact your Leica Representative.
Autofocus not found
• Cable Connection loose.
• Cable damaged.
• Controller defective.
X
Contact your Leica Representative.
Camera control unit not found
The camera may not have been switched on.
X
Switch on the camera.
Contact your Leica Representative if error
persists.
Luxmeter defective error
Photodiode in the optics carrier may be defective
or the light guide is not fully plugged in.
X
Plug in the light guide properly or contact your
Leica Representative if error persists.