8. Operation
Note:
If one of the buttons for changing the operation
mode (DRY or IMM) is pressed by mistake, the
previous mode can be reactivated by pressing
the correct button. The stage will not be
lowered again. The button for the current
objective, however, will remain gray and has to
be pressed again.
Changing the operation mode via:
LeicaScreen - navigation item
Variable function keys at the stand and at
the SmartMove
Software Leica DMControl, module:
DMOperation
Note:
When retrofitting objectives, they have to be
learned in with the Software Leica DMControl,
module: configuration. Afterwards, parfocality
should also be learned in again.
Note:
For lockable immersion objectives lock these by
pushing the front part upwards until it stops
(approx. 2 mm). Then, after a gentle turning
motion to the right, the objective is locked
(Fig. 71).
For objectives with corrective mounts turn the
knurl to adjust the objective to the thickness of
the cover glass.
Fig. 70
Immersion objective (released)
Fig. 71
Immersion objective (locked)
↔↔↔↔↔
↔↔↔↔↔