25
6. Assembly
107/2 Lamp Housing
This lamp housing is used with a 12V 100W halo-
gen lamp
,
which is already mounted.
In case the lamp has to be removed:
• Remove the fastener screw on the housing
(Fig. 15).
• Remove the housing by pulling it upwards.
• Remove the lamp.
• Insert the new 12V 100W lamp (16.1) with the
dust cover straight into the socket until it
stops. Be sure that the lamp is inserted
straight.
• Remove the lamp’s dust cover.
Caution!
Do not remove the lamp’s dust cover until
you have installed the lamp. Avoid
fingerprints on the lamp.
• Replace the housing and fasten it in place us-
ing the fastening screw.
• Place the lamp housing in the transmitted
light lamp housing receptacle (17.2) and fas-
ten it with the clamping screw on the side.
• Connect the lamp housing to the power supply
for transmitted light (symbol: ) (17.3).
• Now connect the lamp power cable of the
microscope (17.5) to the Leica CTR6000
Electronics Box (18.1).
Fig. 17
Rear side of stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol: )
4
12 V 100 W connection for incident light (symbol: )
5
Lamp power cable
1
2
3 4
5
Fig. 18
Rear side of Leica CTR6000
1
Connection for lamp power cable from the stand
1