27
6. Assembly
Fig. 21
Lamp mount with 12 V 100 W halogen lamp
Inserting the 12V 100W halogen lamp into the
106 z lamp housing
• Unscrew the fastening screws of the cover
and lift up the cover (20.1).
• Unscrew the fastening screws of the lamp
mount (20.8) and pull out the mount (Fig. 21).
• Insert the lamp with the dust cover straight
into the socket until it stops.
Caution!
Do not remove the lamp’s dust cover until
you have installed the lamp. Avoid
fingerprints on the lamp.
• Remove the dust cover.
• Reinsert the lamp mount and retighten the
fastening screw (20.8).
• Close the lamp housing and retighten the fas-
tening screws.
• Place the lamp housing in the incident light
lamp housing receptacle (22.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the power supply
for incident light (symbol
) (22.4).
• Now connect the lamp power cable of the
microscope (22.5) to the Leica CTR6000
electronics box (18.1, p. 25)
Fig. 22
Rear side of stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol: )
4
12 V 100 W connection for incident light (symbol: )
5
Lamp power cable
1
2
3 4
5