57
ASP6025 Tissue Processor
Touch the button that reads
TOUCH HERE TO ENTER . . .
in the lower right hand corner during initializa-
tion (
Fig. 58
). This takes you to the
SYSTEM
DIAGNOSTICS
menu (
Fig. 60
).
This menu provides access to basic instrument
settings.
4. Instrument Setup
Fig. 60
1
- Shows the
INSTALLATION
menu (see also
Chapter 5.1.1
).
2 -
Saves the current instrument status to a USB stick.
3 -
Access for service technicians only, the corresponding pass-
word is required.
4 -
Aborts the current program.
5 -
Displays the
RUN LOG
.
3
4
2
1
5
8
7
6
The SYSTEM DIAGNOSTICS menu
The following functions can be selected by touching the corresponding key:
6
- Aborts the current program and deletes the current allocation
of reagents to bottles and retort.
7
- Deletes all reagents and resets the programs and instrument
status.
Caution!
All lists are then deleted.
8
- Restarts the instrument.
Caution!
Only experienced operators may adjust the settings in this menu, since incorrect use of the functions can result
in serious malfunctions.
To exit this menu, the device must be restarted.
To do so, press the RESTART APPLICATION button (8 in
Fig. 60
) and confirm the following prompt with YES. The
initialization restarts as shown in
Fig. 58
and
Fig. 59
.