40
Instructions for Use V 2.4 RevL - 12/2015
4.
Instrument Setup
Fig. 26
4.4 Connecting the power supply
37
39
36
Electrical ports on the back panel of the instrument.
Nameplate for 120 V voltage variant.
Fig. 25
32
33
Adhesive label with fuse ratings
Check the nameplate (
Fig. 25
) on the rear of the
instrument to ensure that the instrument deliv-
ered is designed for the required voltage.
Severe damage may occur if the instrument
is connected to a power supply voltage other
than that to which it was originally set.
The power supply voltage for the instrument
is factory preset and CANNOT be altered by the
user.
Caution!
Observe the following instructions carefully to prevent damage to the instrument:
The 120 V voltage version of the instrument (REF 14 0495 43268) requires an electrical power supply with fuse
protection of at least 20 A.
The instrument MUST be connected to a grounded power socket. The power plug must be easily accessible so
that it can be pulled out.
The instrument is supplied with a set of different power cords. Use only the power cable that matches the local
power supply (outlet).
Do not use an extension cable!