46
Instructions for Use V 2.4 RevL - 12/2015
Fig. 39
• The paraffin station must be pulled out to be filled. It has a cover (
55
) for
better thermal insulation and as splash protection. The cover has two holes
in it to make it easier to remove.
53
4.4.5 The paraffin station
• The paraffin station (
54
) is located behind the left door (
53
) of the instru-
ment, below the retort (
Fig. 37
). It ensures that sufficient fresh, liquid
paraffin is always available for the paraffin baths. The capacity is 5.0 l of
liquid paraffin.
• There are two marks in the interior of the station that display the minimum
fill level when filled with pellets or liquid paraffin (
Fig. 38
). The level must
not fall below these marks.
56
55
4.
Instrument Setup
Fig. 37
56
Enlarged detail:
Sieve in the
Paraffin station
54
Fig. 38
Top mark:
Minimum fill level when filling
with pellets for melting.
Bottom mark:
Minimum fill level when filling
with liquid paraffin.
Fig. 40
• The paraffin station can be filled with paraffin pellets or liquid paraffin. If
filled with pellets, the melting time is approx. 6.0 h.
• Check and clean the strainer (
56
) weekly.
If the paraffin station has been pulled out, this is indicated
by a red border (
Fig. 40
) on the SMART screen.
The signal light on the lower right of the icon will also be
red.
Proceed slowly and carefully when pulling out the station
– never jerk it out. The paraffin inside it is fluid and hot –
it can cause burns. The cover is also hot, therefore always
wear gloves!