178
Instructions for Use V 2.4 RevL - 12/2015
7. Troubleshooting
7.4.2 Mechanical emergency unlock
Only use the mechanical emergency release function if the standard emergency unlock (see
Chap. 7.4.1
) did not work and you must get to the specimens.
After activating the mechanical emergency unlock, working with the instru-
ment is no longer possible!
To trigger the mechanical emergency unlock, follow these steps:
1. Puncture the seal (
Fig. 248, 2
) with the Allen key (
Fig. 248, 1
, included in standard scope of
delivery) and push the slider behind the seal downward.
- Manual unlocking of the retort is triggered.
2. Grasp the bolt handle of the retort firmly in your hand (
Fig. 249
). Since the retort may be under
pressure, open the bolt handle very carefully and slowly.
3. Contact an authorized Leica service technician immediately and notify this person of the
mechanical emergency unlock activation. If you fail to do this, any warranty becomes void.
4. Do not put the instrument back into operation until the inspection is complete.
Fig. 249
Fig. 248
2
1