13
HistoCore AUTOCUT
Safety
2
Warning
Connecting the instrument to a non-grounded socket, or use an extension cord.
Risk of electrical shock which causes injury to persons, or delayed diagnosis.
• The instrument must be connected to a grounded socket.
• Do not use an extension cord.
Caution
Accessories/components loosening/damaged during transportation.
Property damage or delayed diagnosis.
• The packaging has two indicators, ShockDot Impact Indicator and Tilt Indicator, that indicate
improper transport. When the instrument is delivered, check these first. If one of the indicators
is triggered, the package was not handled as prescribed. In this case, please mark the shipping
documents accordingly and check the shipment for damage.
2.2.3 Operating the instrument
Danger
Explosive hazard.
Death or Serious injury of persons, and/or property damage.
• Never operate the instrument in rooms with an explosive hazard.
Warning
Personnel with insufficient qualification working with the instrument.
Serious injury of persons and/or specimen damage can occur when specimen approaches knife/
blade due to inappropriate operator actions, for example, because object head may drop onto knife
holder when handwheel is unlocked.
• Always make sure that only laboratory personnel with dedicated and sufficient training and
qualification operates the instrument.
• Always make sure that all laboratory personnel designated to operate this instrument has read
these Instructions for Use carefully and is familiar with all technical features of the instrument
before start working with it.
Warning
Protective devices or safety accessories provided by manufacturer have been removed or modified.
Serious injury of persons and/or property damage including specimen damage.
• Never remove or modify any protective devices of the instrument or any accessories. Only service
personnel qualified by Leica Biosystems may repair the instrument and access the instrument's
internal components.
• Before working with the instrument always assure that all protective devices and safety
accessories are in place and fulfill their purpose properly.