8
Alarm limit values
Setting a limit value for the cover alarm delay
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A6,
set the required limit value, and save. A timer starts when the cover is opened.
If the timer exceeds the preset delay time, the door alarm is triggered.
Overtemperature limit value
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A13,
set the required limit value, and save. If the refrigeration compartment temperature exceeds the upper preset limit value,
the overtemperature alarm triggers.
Low temperature limit value
Tap the Menu symbol on the touch screen unit. In the parameter level, scroll down to the USr user level, tap parameter A15,
set the required limit value, and save. If the refrigeration compartment temperature falls below the lower preset limit value,
the low temperature alarm triggers.
USr User level
A 6
= Alarm delay (door open)
Factory setting
60 sec.
A 13 = Lower limit 1 (abs/rel)
Factory setting
-4.0 K
A 15 = Upper limit 1 (abs/rel)
Factory setting
4.0 K
C 11 = Refrigeration compartment temperature set point
The value is stored on the device card
C 25 = Hysteresis sensor F1
Factory setting
The value is stored on the device card
H 11 = Offset correction sensor F1
Factory setting
The value is stored on the device card
The device card is located behind the right side cover of the machine compartment of the VF 200.. C and below the up-
per cover of the machine compartment of the VF 550.. C and VF 750.. C.
Содержание VF 20040 C
Страница 2: ...2 42 Versafreeze Chest Freezers V1...
Страница 4: ...4 42 Versafreeze Chest Freezers V1...
Страница 27: ...V1 Versafreeze Chest Freezers 27 42 VF 20040 C 115V_60Hz...
Страница 28: ...VF 20040 C 230V_50Hz...
Страница 29: ...V1 Versafreeze Chest Freezers 29 42 VF 20085 C 115V_60Hz...
Страница 30: ...VF 20085 C 230V_50Hz...
Страница 31: ...V1 Versafreeze Chest Freezers 31 42 VF 55040 C_VF75040 C 115V_60Hz...
Страница 32: ...VF 55040 C_VF75040 C 230V_50Hz...
Страница 33: ...V1 Versafreeze Chest Freezers 33 42 VF 55085 C_VF75085 C 115V_60Hz...
Страница 34: ...VF 55085 C_VF75085 C 230V_50Hz...
Страница 35: ...V1 Versafreeze Chest Freezers 35 42 Touch screen unit Single board cryostat A1...
Страница 38: ...14 Notes...
Страница 40: ...16 Product Returns and Clearance Declaration...
Страница 41: ...V1 Versafreeze Chest Freezers 41 42 17 EC Declaration of Conformity...