V1
Versafreeze Chest Freezers
19 / 42
Water cooling equipment option
Installation of a heat exchanger (water refrigerant) instead of a condenser. The heat exchanger significantly reduces the
heat emitted by the device into the surrounding air, while expanding the permissible working temperature range.
The temperature is pressure-controlled by the freezer appliance. Suitable for connection to a recooling system or water
supply (external thread connection for pressure hose ½
″
with union nut). Other connection types on request.
Not suitable for the models VF 200xx C.
Inlet water pressure min. 0.2 bar, max. 10 bar.
Inlet water temperature min. 4 °C, max. approx. 25 °C.