8.3 Low level alarm
n
An alarm signal sounds if the liquid level falls below the minimum level.
n
Low Level
appears in the display. The components of the device, e.g. the
pump, are switched off via the electronics system.
Personnel:
n
Operating personnel
1.
Top up the missing heat transfer liquid.
2.
Unlock the display with the Enter button.
The device restarts.
8.4 Overtemperature alarm
If the temperature exceeds 90 °C,
3 overtemperature
alarm is triggered and
a two-tone acoustic signal sounds.
The heaters, cooling system and pump are switched off as a result of this
alarm.
1.
Switch off the device.
2.
Allow the device to cool down (> 20 minutes).
3.
Switch on the device.
4.
Press the
OK
key to acknowledge the alarm.
Fig. 68: Low Level Alarm
Rectifying a fault
Fig. 69: Overtemperature alarm
V08
Variocool
74 / 97