6.12.5
Limiting the current consumption
If your mains fuse is less than 16 A, the current consumption can be reduced
in steps from 16 A to 8 A. The output of the heater is reduced in the case
of a device with a heater. This may impair the control accuracy. Establish
whether other consuming units or only your device is connected to the fused
circuit.
Personnel:
n
Operating personnel
1.
Switch to the main menu.
2.
Select the
Settings
Basic setting
Current consumption
menu item.
3.
Change the current consumption accordingly.
4.
Press the input button to confirm.
6.12.6
Configuring the alarm level for the fill level
A warning about low level of the device is usually output on the device
starting from the second level stage. However, the alarm level before low
level can be configured within a specific range.
Personnel:
n
Operating personnel
1.
Switch to the main menu.
2.
Select the
Setup
Basic setup
Level warn stage
menu item.
3.
You can select from four stages 0 to 3 for the warning before low
level. With
3
, a warning about low level is output from the third level
stage. With
0
, no warning at all is output. In this case, the device is
switched off and an alarm displayed when low level is reached.
4.
Press the input button to confirm.
Fig. 43: Specifying current consumption
Fig. 44: Specifying the Alarm level