11.5
Heater (potential equipment)
Table 24: Heating power and power consumption
Alternating current
VC 1200 (W)
VC 2000 (W)
VC 3000 (W)
Unit
230 V; 50 Hz
Heating output
1.5
1.5
1.5
kW
Power consumption with a heater
2.6
2.6
2.6
kW
200 V; 50/60 Hz
Heating output
1.1
1.1
1.1
kW
Power consumption with a heater
2.3
2.3
2.6
kW
208-220 V; 60 Hz
Heating output
1.2 – 1.35
1.2 – 1.35
1.2 – 1.35
kW
Power consumption with a heater
2.4
2.5
2.8
kW
Table 25: Heating power and power consumption
Three-phase current
VC 5000 (W)
VC 7000 (W)
VC 10000 (W)
Unit
400 V; 3/N/PE~50 Hz
Heating output
4.5
4.5
7.5
kW
Power consumption with a heater
7.8
8.8
11.1
kW
208-220 V; 3/PE~60 Hz
Heating output
3.65 – 4.1
3.65 – 4.1
6.1 – 6.9
kW
Power consumption with a heater
4.5
5.7
7.7
kW
200 V; 3/PE~50/60 Hz
Heating output
3.4
3.4
5.7
kW
Power consumption with a heater
4.3
5.4
7.6
kW
Table 26: Boosted heater
Alternating current
VC 1200 (W)
VC 2000 (W)
Unit
230 V; 50 Hz
Heating output
2.25
2.25
kW
200 V; 50/60 Hz
Heating output
1.7
1.7
kW
208-220 V; 60 Hz
Heating output
1.8 – 2.1
1.8 – 2.1
kW
V08
Variocool
88 / 97