Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
89
Parameter
Sl
ave controller (P-controller):
−
Sl
proportional range:
hO
=
Xpf
in
kelvin.
−
Tve, Tde, Prop_E changeover to Auto/ Man
D
@
=
automatic or
D
LM
= manual.
Works setting is automatic. These three parameters should only be
modified by experienced control technicians!
Hold
pressed for 3 seconds, then make changes.
−
Correction
limitation
Dr/
l
See introduction.
Command
Control
Control Parameters
Control
Parameters
−
Open the device parameter menu via the
soft key
Menu
.
−
With the cursor keys continue to
Æ
Control
Æ
Control Parameters
Æ
Control Parameters
.
−
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
−
Where applicable change parameters
marked with (auto) to manual input with
Tv manual/auto
.
−
Select the parameters to be changed with
and confirm with
.
−
Then change the value in the following
settings window and confirm with
.
−
Correction limitation
see introduction.
Help Menu
End T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Correction limitation
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto) 8,3
Xpf
4,0
Prop_E(a) 30
Dr/
l
°C
D @
°C