Proline Low-temperature Thermostats
102
Starting up
YACE0072 / 21.08.07
Example:
Transfer of setpoint 30.5°C to the thermostat with address 15.
Computer
Thermostat
“A015_OUT_SP_00_30.5“CR
Ö
Õ
“A015_OK“CR
7.15.5 Write commands (Data commands to the thermostat)
Command Explanation
OUT_PV_05_XXX.XX
External temperature to be set through the interface.
OUT_SP_00_XXX.XX
Setpoint transfer with up to 3 places before the decimal point and up to 2 places
behind.
OUT_SP_01_XXX
Pump output step 1 to 8.
OUT_SP_02_XXX
Operation mode cooling (0 = OFF / 1 = ON / 2 = AUTOMATIC).
OUT_SP_04_XXX
TiH outflow temperature high limit.
OUT_SP_05_XXX
TiL outflow temperature low limit.
OUT_PAR_00_XXX.X
Setting of control parameter Xp.
OUT_PAR_01_XXX
Setting of control parameter Tn (5...180s; 181 = Off).
OUT_PAR_02_XXX
Setting of control parameter Tv.
OUT_PAR_03_XXX.X
Setting of control parameter Td.
OUT_PAR_04_XXX.XX
Setting of control parameter KpE.
OUT_PAR_05_XXX
Setting of control parameter TnE (0...998s; 999 = Off).
OUT_PAR_06_XXX
Setting of control parameter TvE.
OUT_PAR_07_XXX.X
Setting of control parameter TdE.
OUT_PAR_09_XXX.X
Setting of the max. outflow temperature limit.
OUT_PAR_10_XXX.X
Setting of control parameter XpF.
OUT_PAR_11_XXX
Setting of control parameter TnF (5...180s; 181 = Off).
OUT_PAR_12_XXX
Setting of control parameter TvF.
OUT_PAR_13_XXX.X
Setting of control parameter TdF.
OUT_PAR_14_XXX.X
Setting of the setpoint offset.
OUT_MODE_00_X
Keys Master: 0 = free / 1 = inhibited (corresponds to “KEY”).
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external Analogue / 3 = external
Serial.
OUT_MODE_03_X
Keys Command: 0 = free / 1 = inhibited.
OUT_MODE_04_X
Setpoint offset source: 0=normal / 1=ext.Pt / 2=ext.analog / 3=ext.serial.
START
Switches the unit on (after Standby).
See safety information
Ì
STOP
Switches the unit into Standby (pump, heater, cooling unit OFF).
RMP_SELECT_X
Selection of the program (1...5) to which the further instructions apply. When the
unit is switched on, program 5 is selected automatically.
RMP_START
Start the programmer.
RMP_PAUSE
Hold (pause) the programmer.
RMP_CONT
Restart the programmer after pause.
RMP_STOP
Terminate the program.
RMP_RESET
Delete the program (all Segments).
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_
X
Set a programmer segment (temperature, time, tolerance and pump level). A
segment is added and appropriate values are applied to it.
RMP_OUT_02_XXX
Number of times the program runs: 0 = unlimited / 1...250.
RMP_OUT_06_XXX.XX
Programmer tolerance setting (0 = off / 0.01°C...450.00°C). All following