Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
71
7.8.10 Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
cDE D
3x
to
LMDls
2x
to
B@K--
3x
−
Call the offset works setting for the external temperature probe
cDE D
.
−
Continue as described in
Ì
7.8.8 for the internal temperature
probe.
Command
Default
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to
Æ
Settings
Æ
Calibration
Æ
extern Pt100
Æ
Default
.
−
The adjacent window appears.
−
Confirm selection with
.
−
Continue as described in 7.8.8 for the
internal temperature probe.
Help Menu
End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default