If the Output Monitor terminals are unused a wire link must be fitted between the ‘A’ terminals, and
a second link between the ‘B’ terminals.
Electro-magnetic Door Locks
Use only fail safe door locks provided by Lasermet (electric door strikes or maglocks). These will
prevent access to the room while the laser is on while always allowing people to enter or leave the
room in the event of a power loss. In order to ensure that people can always enter or leave the room
in the event of an emergency it will be necessary to put an emergency stop or break glass switch
near each door. If in doubt call Lasermet technical help.
The door should be fitted with interlock switches (see section 6.2) so that the laser is disabled if a
door is opened using the break glass switch.
Note that several maglocks can be wired in parallel from the same ICS-7-OEM output provided the
4A maximum rating of the contact is not exceeded.