Section #8 – Troubleshooting Tips
53
Electronic Copy of LTI's UltraLyte LR B User's Manual - 2
nd
Edition © 2002
Section #8 – Troubleshooting Tips
What You Will See
Required Action
Unit powers OFF
by itself.
-or-
No power at all.
•
Verify that the batteries are installed
correctly.
•
Replace the batteries.
•
Keep in mind that the instrument
automatically powers OFF if there is
no instrument activity for a period of
15 minutes.
The in-scope red aiming
dot is not visible.
•
Press the
TRIGGER
to activate
the red dot.
•
Press the
BRIGHTNESS CONTROL
button
to increase the intensity of the
red dot.
•
Adjust the polarizing filter.
E01 error code. No
range or speed
readings.
•
Can you measure to a wall that is
about 50 feet (or 20 meters) away?
•
Check the scope alignment.
•
Restore the factory defaults.
•
When measuring a short range to a
small target, aim slightly above the
target.
Erroneous readings/not
repeatable.
•
Do you have a clear line of sight?
•
Check the scope alignment.
•
Is it raining or foggy. If yes, is the
weather filter active? If not, turn it
ON. Keep in mind that the minimum
range will be 250 feet (76 meters).
•
When measuring a short range to a
small target, aim slightly above the
target
No serial
communication.
•
Verify the cable connection.
Содержание UltraLyte LR B
Страница 1: ...UltraLyte LR B User s Manual...
Страница 6: ...Safety Precautions 4 Electronic Copy of LTI s UltraLyte LR B User s Manual 2nd Edition 2002...
Страница 52: ...Section 6 Serial Data Interface 50 Electronic Copy of LTI s UltraLyte LR B User s Manual 2nd Edition 2002...
Страница 54: ...Section 7 Specifications 52 Electronic Copy of LTI s UltraLyte LR B User s Manual 2nd Edition 2002...