LTI TruPoint 200h User’s Manual • Section 3 - Basic Operations
Page 23
Copyright (c)[2020] Laser Technology, Inc. All Rights Reserved. Unauthorized duplication, in whole or in part is strictly prohibited.
Error Codes
Error conditions can occur in a measurement or
in the system hardware. To make sure that you never
get an erroneous measurement, the TruPoint 200h
monitors both system hardware and measurements.
Error codes appear in the main display and are in
the form of “E xx”, where “xx” is an error code
number (Figure 16).
Error Code
Explanation
Error 02 (E 02)
Insufficient
The user released the fire button before the instrument
acquired/validated a target.
Error 03 (E 03)
Unstable
The return signal from the target is varying too much.
Error 06 (E 06)
Tilt Error
Tilt data out of expected range.
Error 09 (E 09)
Rejected
The acquired target is not in the user Range Gate values/settings.
Error 52 (E 52)
Too Cold
The temperature is too low, displayed when the user tries to start
a measurement.
Error 53 (E 53)
Too Hot
The temperature is too high, displayed when the user tries to start
a measurement.
Error 54 (E 54)
Battery
The battery voltage is too low displayed when the user tries to start
a measurement.
Error 55 (E 55)
or above
Shut off unit and try again. If the same error repeats,
please contact LTI Service Department.
If an error code persists:
1.
Release the
button and press again to try to retake the
measurement.
2.
Remove and re-install the battery and then try to retake the
measurement.
3.
If the above steps do not resolve the error, contact LTI or an
Authorized LTI Distributor for assistance. Refer to the inside
front cover for LTI contact information.
Figure 16
**Draft B**