4-4
Measurement Settings Tabs
LxT Manual
SLM Tab
Figure 4-2 shows the
SLM
tab.
FIGURE 4-3 SLM Tab
To modify settings on the
SLM
tab, navigate to either
Frequency
Weighting
,
Detector
,
Peak Weighting
, or
Integration Method
page and press
5
.
Frequency Weighting
A, C and Z frequency weightings are provided for the RMS
and peak detectors. These are selected separately.
Integration Method
Two Integration methods are available:
Linear
and
Exponential
.
Exponential Integration
Exponential integration would typically be selected to
provide compatibility with older analog instruments in
which measurements are exponential time weighted signals.
Exponential detectors tend to hide small events in the long
decay of loud impulsive events.
Linear Integration
Linear integration utilizes sampled sound pressure levels to
compute RMS levels directly, without an intermediate time
weighting.
Содержание SoundExpert LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...