LxT Manual
Technical Specifications
A
-
59
Vibration Sensitivity
In these tests, the LxT is mounted on an electrodynamic exciter and vibrated sinusoidally at an
amplitude of 1.0 m/s
2
at the following frequencies: 35.5, 63, 125, 250, 500, 630, 800 and 1, 000 Hz.
The tests are performed with the meter oriented such that vibrations are produced in two different
directions: parallel to the microphone diaphragm and perpendicular to the diaphragm.
The resulting sound levels are measured using two different microphones:
The microphone connected to the instrument, which is moving
A reference microphone at a fixed position
The following two sections show diagrams of the test setup and the data measured.
Vibration Parallel to the Microphone Diaphragm
FIGURE A-29 Vibration Parallel to Microphone Diaphragm
Frequency
L
aeq
Ref
31.5
35.7
35.4
63
35.5
34.8
125
35.8
36.8
250
47.7
47.9
Table A-29 Axis of Vibration Parallel to the Microphone Diaphragm
REFERENCE MICROPHONE
VIBRATION PARALLEL
TO MICROPHONE DIAPHRAGM
Vibration
Содержание SoundExpert LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...