LxT Manual
Sound Exposure Metrics Measured
16-3
Spectral data is displayed on both the
Live
and
Overall
tabs,
but only the Overall Data can be stored.
Live Tab
From the
Live
tab, the graphic shows the instantaneous SPL
value for all frequencies and the bar to the far right shows
the summation value for the entire frequency band. The
value corresponding to the cursor position is displayed
numerically beneath the graph.
Overall Tab
From the
Overall
tab, the graphic shows the energy
equivalent level calculated over the measurement time
period at each frequency band and, at the far right, for the
summation of all frequency bands. The values displayed
digitally beneath the graph represent the following data for
the frequency band at the cursor position.
• Leq
• Lmax
• Lmin
Sound Exposure Metrics Measured
The LxT measures two separate and independent sets of
sound exposure metrics.
The following parameters are user-selectable:
• Exchange Rate: 3, 4, 5 or 6 dB
• Threshold Enable: Yes or No
• Threshold Level: Numeric entry
• Criterion, Level and Hours: Numeric entries
Содержание SoundExpert LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...