TF1100-EH transit-time handheld flowmeter manual
25
cords are
not
properly
connected
J
Hardware Error
Hardware
problem
Contact the factory
H
PoorSig Detected
(1)Poor signal
detected
(2)Transducers
installed
improperly
(3)Too much
fouling
(4)The pipe
liners are
too thick.
(5)Problem
with
transducers
cords
(1)Relocate measuring place
(2)Clean the spot
(3)Check the cords
(4)Check the coupler
Q
Freq. Output Over
The actual
frequency for
the Frequency
Output is out
of the range
set by the user
Check the value entered at M66,M67,M68
and M69, and try to enter a larger value on
M69
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
(1) Temporary
problems
with
RAM,
RTC
(2) Permanent
problems
with
hardware
(1) power on again
(2) contact factory
1
2
3
Adjusting Gain
Instrument is
in the progress
of adjusting
the gain for
the signal, and
the number
indicates the
progressive
steps
K
Empty pipe
(1) No liquid
inside pipe
(2)
Setup error
on M29
(1)
Relocate where the pipe is full of liquid
(2)
Enter 0 on M29