2.5 STEPS TO CONFIGURE THE PARAMETERS
The following parameters need to be configured for a proper measurement:
(1)
Pipe outer diameter (O.D.)
(2)
Pipe wall thickness
(3)
Pipe materials (for non-standard pipe materials*, the sound speed for the material must be
configured too)
*Standard pipe materials and standard liquids refer to those with the sound parameters
that have already been programmed into software of the flow meter; therefore there is no
need to configure them.
(4)
Liner material and its sound speed and thickness, if there is any liner.
(5)
Liquid type (for non-standard liquids, the sound speed of the liquid is also needed)
(6)
Transducer type adapted to the flow meter. Generally the Standard M1 or the Frame
M-sized transducers will be the selected option.
(7)
Transducer mounting methods (the V-method or Z-method is the common option)
(8)
Check up the Space displayed on M25 and install the transducers accordingly.
For standard pipe materials and standard liquids, the following detailed step-by-step setup is
recommended.
(1) Press keys MENU 1 1 to enter M11 window to input the digits for the pipe outer
diameter, and then press ENT key.
(2) Press key
∨
/- to enter M12 window to input the digits for the wall thickness and then
press ENT key.
(3) Press key
∨
/- to enter M14 window, and press ENT key to enter the pipe material.
Use keys
∧
/+ and
∨
/- to scroll up and down to the intended pipe material, and then
press ENT key. If you select “Other”, you must enter Menu15, pipe sound speed.
(4) Press key
∨
/- to enter M16 window, press ENT key to enter the option selection
mode, use keys
∧
/+ and
∨
/- to scroll up and down to the liner material, and then
press ENT key. Select “No Liner”, if there is no liner, If you select “Other”, you must
enter Menu17, liner sound speed.
(5) Enter “Liner Thickness” in Menu18.
(5) Press key
∨
/- to enter M20 window, press ENT key to enter Fluid Type, use keys
∧
/+ and
∨
/- to scroll up and down to the proper liquid, and then press ENT key.
If you select “Other”, you must enter Menu21, Fluid sound speed and Menu 22, Fluid
Viscosity
(6) Press key
∨
/- to enter M23 window, press ENT key to enter the Transducer Type,
use keys
∧
/+ and
∨
/- to scroll up and down to the proper Transducer Type (Standard
S, M, L) you selected, and then press ENT key. Plug Type C and B are for Insertion
transducer, no matter with Clamp-on transducers.
(7) Press key
∨
/- to enter M24 window to select Transducer Mounting mode, and then