Appendix 2
Error Codes
At switch-on, the
Cyclops
instrument will check all stored data in its non-
volatile memory and any detected errors will be shown near the bottom of the
side display screen.
Error Code 1
Description:
Unknown Cyclops Type
If this code is displayed the instrument will be unable to perform any further
any function.
If the error persists, return the
Cyclops
to Land Instruments for repair.
Error Code 2
Description:
Lost Calibration Data
The
Cyclops
will continue to work, but with the display inverted (i.e. white
on black). However, the calibration of the instrument will be incorrect and
readings will only be approximate. It will not meet the temperature accuracy
specifications. If this error persists, return the
Cyclops
to Land Instruments
for repair.
Error Code 3
Description:
Lost User Setting
The
Cyclops
normally stores user settings (accessed via the menu) in non-
volatile memory. If these are inconsistent, the user settings will revert to
factory defaults. The operator can re-enter the desired user settings to allow
the
Cyclops
to work as intended.