9. Turn the bottle right side up and remove the
Titrator.
10. Insert the adapter tip into the opening of the tiration
tube cap. Slowly depress the
plunger to dispanse the titrating solution. Gently
swirl the tube to mix the solution.
11. Continue adding the titration solution until the
speci
fi
ed color change occurs. If the color change
does not occur when the large ring of the plunger
reached the bottom of the scale, re
fi
ll the Titrator to
the zero line. Continue titration until the color change
occurs.
12. Read the result directly from the scale where the
large ring on the Titrator meets the Titrator barrel.
If the Titrator was re
fi
lled to reach the
fi
nal color
change, add the total amounts of titrant used to
determine the
fi
nal result.
13. If no additional, tests are to be made, discard the remaining titration
solution in the Titrator. Do not return the titrating solution to the reagent
bottle. Thoroughly rinse the titration tube and the Titrator.
Do not remove
the plunger or the adapter tip from the Titrator
.