LAMBDA OMNICOLL Fraction Collector - Sampler
Operation MANUAL 2023.01
www.lambda-instruments.com
24
You can change this value by means of the four buttons
Λ Λ Λ Λ
under the display. Press the
button
SET
to confirm the new value (long beep).
Figure 3.7-4
Selecting parameter
COUNT
(volume).
[Count LED: ON;
Time LED: OFF;
Display: coUn]
Figure 3.7-5
Selecting the parameter
TIME
to
perform the program for a desired time.
[Count LED: OFF;
Time LED: ON;
Display: tiMe]
Figure 3.7-6
Selecting the parameter
PAUSE
for introducing pauses in the program.
[Count LED: ON;
Time LED: ON;
Display: PAUS]
Figure 3.7-7
Selecting the parameter
NUMBER
for collecting a number of samples.
[Count LED: OFF;
Time LED: OFF;
Display: nUMb]
After setting of the parameters, the fraction collection is started by pressing the
START
button.
The collector will take the number of samples as programmed under the
NUMBER
setting
and will stop for the time programmed under the
PAUSE
setting. This cycle will go on until a
STOP signal will be encountered. The LAMBDA peristaltic pump is activated automatically
only during fraction collection.
When the fraction collection is controlled from the panel (i.e., not by remote
control signal) the fraction collector will collect the number of
fractions/samples introduced in the parameter
NUMB
of duration
TIME
or
volume
COUNT
(if a drop counter or LAMBDA pumps are used), followed
by the PAUSE, if a PAUSE was set.
Then, the fraction collector will repeat this till the
STOP
signal is
encountered (i.e., STOP is pressed or the fraction collector is stopped by
the end magnet). This collection mode allows for e.g., a line washing step
before the relevant fraction is taken.