LAMBDA OMNICOLL Fraction Collector - Sampler
Operation MANUAL 2023.01
www.lambda-instruments.com
19
A small difference in the fraction positions with respect to the tube centres
may occur between fractions taken in opposite collection directions. If the
fraction lines are not too broad (about 2 mm), this difference in position is of
approx. 1 mm
3.4.2 Time resolution (0.1 or 1 minute)
The time resolution of the OMNICOLL fraction collector can be selected in the following way:
Figure 3.4-6
Press the button
COUNT/TIME
while connecting the power supply to the mains.
Figure 3.4-7
The actual time resolution will
appear on the display
“0.1M”
or
“1M”.
Figure 3.4-8
By pressing the button
COUNT /
TIME
below the COUNT and TIME LEDs you
can select to use 0.1 minute steps
“0.1M”
or 1
minute steps
“1M”.
Figure 3.4-9
Confirm the choice by pressing the
button
SET
.