Lake Shore Model 480 Fluxmeter User’s Manual
Instrument Setup
3-1
CHAPTER 3
INSTRUMENT SETUP
3.0 GENERAL
This chapter provides setup information for the Lake Shore Model 480 Fluxmeter. A general layout for
the Model 480 rear panel and information on how to make line power, coil, probe, and terminal block
connections to the Model 480 are provided. If you want to experiment with the various software settings
covered in the next chapter before doing a complete hardware setup, the Model 480 may be operated
with only the line power connected (i.e., no coil, probe, or terminal block connections).
CAUTION:
Before plugging in the Model 480 and turning it on,
read about line voltage settings in
Paragraph 3.3. An improper line voltage setting may damage the Model 480. Check it
carefully before powering the instrument for the first time.
This chapter covers receiving the Model 480 in Paragraph 3.1, rear panel control definitions in
Paragraph 3.2, power line input assembly in Paragraph 3.3, coil input connections in Paragraph 3.4,
probe input connection in Paragraph 3.5, and terminal block in Paragraph 3.6.
3.1 RECEIVING THE MODEL 480
This section covers inspection and unpacking in Paragraph 3.1.1 and repackaging for shipment in
Paragraph 3.1.2.
3.1.1 Inspection and Unpacking
Inspect shipping containers for external damage. Make all claims for damage (apparent or concealed)
or partial loss of shipment in writing to Lake Shore within five (5) days from receipt of goods. If
damage or loss is apparent, please notify the shipping agent immediately.
Use the packing list included with the instrument to verify receipt of the instrument, probe and/or coil,
accessories, and manual. Inspect for damage. Inventory all components supplied before discarding
any shipping materials. If there is freight damage to the instrument, file proper claims promptly with
the carrier and insurance company and notify Lake Shore. Notify Lake Shore immediately of any
missing parts. Lake Shore cannot be responsible for any missing parts unless notified within 60 days
of shipment. Refer to the standard Lake Shore Warranty on the A Page (behind the title page).
3.1.2 Repackaging For Shipment
To return the Model 480, probe, coil, or accessories for repair, replacement, or recalibration, obtain a
Return Goods Authorization (RGA) number from Technical Service in the United States, or from the
authorized sales/service representative from which the product was purchased. Instruments may not
be accepted without a RGA number. When returning an instrument for service, Lake Shore must
have the following information before attempting any repair.
1. Instrument model and serial number.
2. User name, company, address, and phone number.
3. Malfunction symptoms.
4. Description of system.
5. Returned Goods Authorization (RGA) number.
Wrap instrument in a protective bag and use original spacers to protect controls. Repack the system
in the Lake Shore shipping carton (if available) and seal it with strong paper or nylon tape. Affix
shipping labels and FRAGILE warnings. Write the RGA number on the outside of the shipping
container or on the packing slip.
Содержание 480
Страница 6: ...Lake Shore Model 480 Fluxmeter User s Manual C This Page Intentionally Left Blank ...
Страница 12: ...Lake Shore Model 480 Fluxmeter User s Manual vi Table of Contents This Page Intentionally Left Blank ...
Страница 66: ...Lake Shore Model 480 Fluxmeter User s Manual 5 24 Advanced Operation This Page Intentionally Left Blank ...
Страница 102: ...Lake Shore Model 480 Fluxmeter User s Manual 6 36 Computer Interface Operation This Page Intentionally Left Blank ...