Lake Shore Model 480 Fluxmeter User’s Manual
8-8
Service and Calibration
8.6 TOP OF ENCLOSURE REMOVAL AND REPLACEMENT
WARNING: To avoid potentially lethal shocks, turn off controller and disconnect it from AC
power line before performing this procedure. Only qualified personnel should
perform this procedure.
8.6.1 Removal Procedure
1. Set power switch to Off (
O
) and disconnect power cord from rear of unit.
2. If attached, remove 19-inch rack mounting brackets.
3. Use 5/64 hex key to remove four screws attaching top panel to unit.
4. Use 5/64 hex key to loosen four screws attaching bottom panel to unit.
5. Carefully remove the back bezel by sliding it straight back away from the unit.
6. Slide the top panel back and remove it from the unit.
8.6.2 Installation Procedure
1. Slide the top panel forward in the track provided on each side of the unit.
2. Carefully replace the back bezel by sliding it straight into the unit.
3. Use 5/64 hex key to install four screws attaching top panel to unit.
4. Use 5/64 hex key to tighten four screws attaching bottom panel to unit.
5. If required, reattach 19-inch rack mounting brackets.
6. Connect power cord to rear of unit and set power switch to On (
l
).
8.7 EPROM REPLACEMENT
The operating software for the Model 480 is contained on one Erasable Programmable Read Only
Memory (EPROM) Integrated Circuit (IC). The reference designator for the EPROM is U53. The
EPROM has a sticker on top labeled with “M480.HEX” and the date. Use the following procedure to
replace the EPROM.
CAUTION
: The EPROM is an Electrostatic Discharge Sensitive (ESDS) device. Wear shock-proof
wrist straps (with a resistor that limits current to <5 mA) to prevent injury to service
personnel and to avoid inducing an Electrostatic Discharge (ESD) into the device. Refer to
Paragraph 8.2.
1. Follow the top of enclosure REMOVAL procedure in Paragraph 8.6.1.
2. Remove four Phillips-head screws attaching transformer bracket to the Model 480 chassis.
Carefully pull transformer bracket up and sufficiently out of the way to gain access to the Operating
Software EPROM. See Figure 8-7.
3. Locate EPROM U53 (M480.HEX) on the main circuit board. Note orientation of existing IC.
eprom.eps
4. Use IC puller to remove existing EPROM from socket.
5. Noting orientation of new EPROM, use an IC insertion tool to place new device into socket.
6. Install four Phillips-head screws attaching transformer bracket to the Model 480 chassis.
7. Follow the top of enclosure INSTALLATION procedure in Paragraph 8.6.2.
Содержание 480
Страница 6: ...Lake Shore Model 480 Fluxmeter User s Manual C This Page Intentionally Left Blank ...
Страница 12: ...Lake Shore Model 480 Fluxmeter User s Manual vi Table of Contents This Page Intentionally Left Blank ...
Страница 66: ...Lake Shore Model 480 Fluxmeter User s Manual 5 24 Advanced Operation This Page Intentionally Left Blank ...
Страница 102: ...Lake Shore Model 480 Fluxmeter User s Manual 6 36 Computer Interface Operation This Page Intentionally Left Blank ...