Lake Shore Model 480 Fluxmeter User’s Manual
Introduction
1-1
CHAPTER 1
INTRODUCTION
1.0 GENERAL
This chapter provides introductory information for the Lake Shore Model 480 Fluxmeter. Product
description is in Paragraph 1.1, specifications in Paragraph 1.2, safety summary in Paragraph 1.3, and
safety symbols in Paragraph 1.4.
1.1 PRODUCT DESCRIPTION
The Model 480 is a precision integrating fluxmeter that works with a variety of sensing coils to measure changing
flux. It is fundamentally an analog integrator under microprocessor control. The analog integrator has excellent
specifications and is very flexible. It performs well in a variety of magnet applications from a fast pulse to a slow
ramp. The microprocessor optimizes the performance of the integrator and enables numerous features and
interfaces. The Model 480 fits well into test and measurement operations from all manual to fully automated with
quick setup and ease of use. The fluxmeter complements the existing line of Lake Shore gaussmeters.
Manual Magnet Testing
A bright display and fast reading update make the Model 480 ideal for manual magnet sorting and testing. The low
drift of the instrument improves productivity with fewer adjustments. Remote terminals allow for foot pedal reading
reset to keep hands on the work, not the instrument. Configurable alarms give an audible signal or relay closure to
signify pass/fail.
Automated Magnet Testing
In automated testing, time is money. The Model 480 has many features to enhance throughput. The instrument has
a fast update rate and settling time. It recovers quickly from reading reset to start a new reading cycle. IEEE-488
and serial computer interfaces included with the Model 480 can be used to control most instrument functions.
Relays and analog outputs can be used for automation without a computer interface.
Magnetizing
The magnetizing process places unique demands on all associated electronics. The Model 480 responds with very
fast peak capture that can keep up with the fastest magnetizing pulses. Both a positive and negative peak can be
captured from the same pulse. The input of the Model 480 is protected against the high voltages at its input present
during magnetizing.
Materials Analysis
High resolution and low drift define the role of the fluxmeter in analytical measurement. The high resolution of the
Model 480 is reinforced by a low noise floor. A configurable filter helps keep the readings quiet. Automatic and
manual drift adjustment modes help optimize the low-drift characteristics of the integrators. The IEEE-488 and
serial computer interfaces included with the Model 480 allow automated data taking.
AC Magnetic Fields
Sensing coils are sensitive to AC magnetic fields but many conventional integrating fluxmeters can not measure
AC fields. The Model 480 has an AC mode that enables it to measure fields over a wide frequency range using
simple sensing coils. Applications are limited to field volumes as large or larger than the coil but for some, it is an
inexpensive way to make low drift AC field measurements.
Drift Adjustment
Adjusting or nulling the drift of an analog integrator wastes time. It can be the only unpleasant part of using an
integrating fluxmeter. Lake Shore innovation brings some relief. The Model 480 has a built in drift algorithm that
continually adjusts drift when the instrument and coil are idle. It is ready when you are to make precision low drift
measurements. The adjustment algorithm has no effect during flux integration. Manual drift adjustment is also
available.
Coils and Probes
Coils and probes wound by the user or from other manufacturers can easily be used with the Model 480. The
Model 480 allows the user to save parameters for up to 10 existing coils/probes and quickly switch between them.
Lake Shore also offers several sensing coils and probe assemblies for use with the Model 480 which offer several
conveniences. They are factory calibrated for accuracy and interchangeability. Calibration data is loaded into
memory in the probe connector so it does not have to be entered by the user. Special coil assemblies are also
available and can be designed to meet customer specifications.
Содержание 480
Страница 6: ...Lake Shore Model 480 Fluxmeter User s Manual C This Page Intentionally Left Blank ...
Страница 12: ...Lake Shore Model 480 Fluxmeter User s Manual vi Table of Contents This Page Intentionally Left Blank ...
Страница 66: ...Lake Shore Model 480 Fluxmeter User s Manual 5 24 Advanced Operation This Page Intentionally Left Blank ...
Страница 102: ...Lake Shore Model 480 Fluxmeter User s Manual 6 36 Computer Interface Operation This Page Intentionally Left Blank ...