Lake Shore Model 460 Gaussmeter User’s Manual
5-14
Accessories & Probes
NOTE:
Use care to ensure the Zero Gauss Chamber does not become magnetized. Using a
magnetized chamber to zero a probe can lead to erroneous field readings. It is a good
practice to periodically degauss the chamber. If no professional degausser is available,
a bulk tape degausser (Verity VS250, Data Devices PF211, or equivalent) may be used.
C-460-5-15.eps
Figure 5-15. Model 4060 Zero Gauss Chamber
C-460-5-16.eps
Figure 5-16. Model 4065 Large Zero Gauss Chamber