Lake Shore Model 460 Gaussmeter User’s Manual
Remote Operation
4-31
Device Specific Commands (Continued)
ALMH
Alarm High Point Command
Input:
ALMH <field value>[term]
Format:
±nnn.nn
<field value>
Returns sign, 4 or 5 digits, and places decimal point appropriate to range.
Remarks:
New value is entered on the same field range as the old value. Setting value to zero first will
change the setting range to present display range.
ALMH?
Alarm High Point Query
Input:
ALMH?[term]
Returned:
<field value>[term]
Format:
±nnn.nn (Refer to command for description)
Remarks:
Use ALMHM? to determine units multiplier.
ALMHM?
Alarm High Point Multiplier Query
Input:
ALMHM?[term]
Returned:
<multiplier>[term]
Format:
a
<multiplier>
u = micro = 10
-6
, m = milli = 10
-3
, (blank) = unity, k = kilo = 10
3
Remarks:
Used with ALMH? query.
ALMIO
Alarm Inside/Outside Command
Input:
ALMIO <inout>[term]
Format:
n
<inout>
0 = Outside or 1 = Inside
Remarks:
Instructs alarm feature to cause an active alarm state when the field reading is either inside of
or outside of the high and low setpoint values. Refer to Paragraph 3.10.
ALMIO?
Alarm Inside/Outside Query
Input:
ALMIO?[term]
Returned:
<inout>[term]
Format:
n
(Refer to command for description)
ALML
Alarm Low Point Command
Input:
ALML <field value>[term]
Format:
±nnn.nn
<field value>
Returns sign, 4 or 5 digits, and places decimal point appropriate to range.
Remarks:
New value is entered on the same field range as the old value. Setting value to zero first will
change the setting range to present display range.