Guidance and manufacturer’s declaration
electromagnetic emissions
The NEBULIZER MODEL NE1005 is intended for use in the electromagnetic environment
specified below. The customer or the user of the NEBULIZER MODEL NE1005 should assure
that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment-
guidance
RF emissions
CISPR 11
Group 1
The NEBULIZER MODEL NE1005
uses RF energy only for its
internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
T h e N E B U L I Z E R M O D E L
NE1005 is suitable for use in
all establishments, including
domestic establishments
and those directly connected to
the public low-voltage
power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration
electromagnetic immunity
The NEBULIZER MODEL NE1005 is intended for use in the electromagnetic environment
specified below. The customer or the user of the NEBULIZER MODEL NE1005 should assure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
E l e c t r o m a g n e t i c
e n v i r o n m e n t
-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines and dc lines
±2 kV for power
supply lines and dc lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) and neutral
±1 kV line(s) and neutral
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations
on power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
for 0,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5s
<5% UT
(>95% dip in UT)
for 0,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles)
<5% UT
(>95% dip in UT)
for 5s
Mains power quality
should be that of a
typical commercial or
hospital environment. If
a dips or an interruption
of mains power occurs,
t h e c u r r e n t o f t h e
NEBULIZER MODEL
NE1005 may be dropped
off from normal level, it
may be necessary to use
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
P o w e r f r e q u e n c y
m a g n e t i c f i e l d s
should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level