Test No.
Test result indication
Action if test failed
Test 1 -
Occlusions
•
Check possible blockages (e.g. twisted tub-
ing, blocked filter, blocked filter in the liner)
and run the Device. Test again.
•
If the High Efficiency Filtration Kit is installed
the pass limit is 150mmHg.
Test 2 – Vacuum
efficacy
•
Check Connectors, Tubes and Canister Lid
for leakage* or damage.
•
Check exhaust outlet for occlusion and run
the Device Test again.
Test 3 – Maxi-
mum vacuum
•
Check Connectors, Tubes and Canister Lid
for leakage* or damage.
•
Check exhaust outlet for occlusion and run
the Device Test again.
Test 4 - Leakages
Check Connectors, Tubes and Canister Lid for
leakage* or damage and run the Device Test
again.
After evaluating the test results turn the Operating Knob to “0” to exit the Device Test.
Evaluation of Device Test Results
After the test is completed, the Vacuum Indicator
will display the results. Press the Test Button to scroll
through the results of each test to display the results.
Device Test
Test Passed
<100 mmHg
Test Passed
>300 mmHg
Test Passed
>500 mmHg
Test Passed
>450 mmHg