13
Caution
Even if the Operating Knob is set to “0”, voltage is present on some of the internal circuitry
when the LSU is connected to external power. Disconnect from the mains to fully remove
power.
Note
The LSU has an automatic power save mode which switches the pump motor off. While in
this mode, the Power On Indicator will flash slowly (approx. once per second). Power save
mode is activated when the Operating Knob is set to 200, 350 or 500+ mmHg and the
actual vacuum level has been continuously higher than 120 mmHg for more than 2 minutes.
To exit power save mode and revert to normal operation, set the Operating Knob to any other
position and then go back to required setting.
After Use Checklist
• Inspect all parts of the LSU for damage and excessive wear. Replace parts if necessary.
• Clean the LSU cabinet. Clean and disinfect reusable parts. See Cleaning section.
• LSU Serres: The Serres Vacuum Connector should be replaced regularly. Place an
unopened Serres Suction Tube on the side of the LSU.
• Perform Device Test. See
Device Test
section.
• Place the LSU on charge.
The Aerosol Filter protects the LSU by preventing aerosols from entering the Pump Unit.
It is not intended for microbiological or for particle filtration. The Aerosol Filter is not
designed for decontamination. It is recommended that the Aerosol Filter be replaced
after each use or at least once every shift. If the LSU is used on patients in areas where
cross contamination is not an issue, the Aerosol Filter should be replaced at least once a
month. It is recommended always to have extra Aerosol Filters with the LSU in case one
has to be discarded. If the Aerosol Filter becomes wet, it should be replaced immediately
or as soon as possible after use.
Note
Check the filter after each use. If the filter is broken so liquid penetrates the membrane, the
pump will be contaminated. Contact Laerdal Service.
Note
The Float Ball (in the Canister lid) shuts off the vacuum if the Canister is full or the LSU tips
over. To restore the vacuum, remove the Angled Connector from the Vacuum Inlet. The Float
Ball will be released and the Angled Connector can be reattached.
Note
The Serres Suction Bag system contains a hydrophilic filter that shuts off the vacuum if the
canister is full or the LSU tips over. To restore the vacuum, replace the Suction bag.