Steam 100 °C
10-20 minutes
Steam 100 °C
10-20 minutes
Steam 100 °C
10-20 minutes
16
Decontaminating and disinfection of Reusable Canister – LSU Reusable
1. Dismantle
• Dismantle parts to be
decontaminated after each use.
• The Float Ball can be snapped out of
the lid.
2. Rinse
• Rinse all parts under cold running
water a total of 3 times.
• Immerse in warm water.
3. Clean
• Immerse all parts in hot water
(60-70 °C) containing a mild
detergent.
• Thoroughly clean all surfaces, use a
brush where possible.
• Rinse in warm water and allow to
dry.
• Inspect all parts to be visibly clean
and dry.
• Note
Thorough rinsing and cleaning
are very important steps prior to
disinfection.
Cleaning
3 X
Содержание Reusable