Ultimate-cell Stackable Shaking Incubator V2017.2
42
Press "CALIB" in the "SET" page to enter the page of calibrations.
In the
“CALIB” page, press “Password” icon, enter “7724”, and press “OK” to
confirm.
Press the
“CALIB TEMP” button, which is showing the current displayed tem-
perature value. Use the keypad to change the temperature value to the cali-
bration value.
When the desired calibration value is displayed, press the "OK" to confirm and
save.
7.2
CALIB HUMI
—Calibrating the Relative Humidity Display
Place the independent humidity reference device in the geometrical center of
the chamber.
If the incubator has just been installed and started up, allow another 30-60mins
once the incubator to stabilize at the setpoint before making a calibration ad-
justment.
Press "SET" in the home page
Press "CALIB" in the "SET" page to enter the page of calibrations.
In the
“CALIB” page, press “Password” icon, enter “7724”, and press “OK” to
confirm.
Press the
“CALIB HUMI” button, which is showing the current displayed relative
humidity value. Use the keypad to change the relative humidity value to the
calibration value.
When the desired calibration value is displayed, press the "OK" to confirm and
save.
7.3
CALIB CO2
—CO2 Calibrating the CO2 Concentration Display
Place the independent CO2 gas concentration reference device in the geo-
metrical center of the chamber.
If the incubator has just been installed and started up, allow 8-12 hours for the
incubator to stabilize at the setpoint before making a calibration adjustment. If
the unit has been in operation, allow a full hour after the display reaches the
CO2 setpoint before making any calibration adjustments.
Press "SET" in the home page
Press "CALIB" in the "SET" page to enter the page of calibrations.
In the
“CALIB” page, press “Password” icon, enter “7724”, and press “OK” to
confirm.
Press the
“CALIB CO2” button, which is showing the current displayed CO2
concentration value. Use the keypad to change the CO2 concentration value to
the calibration value.
When the desired calibration value is displayed, press the "OK" to confirm and
save.