Ultimate-cell Stackable Shaking Incubator V2017.2
41
7. CALIBRATING THE UNIT
After the incubator has stabilized at the user selected setpoints, the control systems can be
calibrated. When in calibration, the temperature display, humidity display, CO2 display, and
can all be adjusted to match a reference standard.
Before any system is calibrated it is critical that the incubator has been allowed to stabilize
properly. In the instructions below, each individual system has a recommended stabilization
time that is required before calibration.
To calibrate each of the these parameters,
7.1
CALIB TEMP
—Calibrating the Temperature Display
Place the independent temperature reference device in the geometrical center
of the chamber.
If the incubator has just been installed and started up, allow another 30-60mins
once the incubator to stabilize at the setpoint before making a calibration ad-
justment.
Press "SET" in the home page