64
© 2008
– 2020 by MMS B.V.
Chapter 4
The measured volume may deviate as follows (when the measured values are
not within the allowed limits, you can clean the catheter or calibrate the
transducers).
Applied pressure
(cmH
2
O)
Allowed limits (cmH
2
O)
50
100
± 2 cmH
2
O ± deviation of catheter
± 2 cmH
2
O ± deviation of catheter
You can also test all transducers connected to the Solar Blue module, a
selection of these transducers, or the complete catheter with all transducers
based on a catheter definition.
Pressure unit
The pressure values can be shown in four pressure units: cmH
2
O, inches H
2
O,
kPa and mmHg. To change the pressure unit, click the pressure unit combo-
box and select another pressure unit.
Below you can find a pressure unit conversion table.
Pressure unit
Conversion
1 mmHg
0,13 kPa
1,36 cmH
2
O
0,54 inchH
2
O
1 cmH
2
O
0,10 kPa
0,74 mmHg
0,39 inchH
2
O
1 kPa
10,20 cmH
2
O
4,01 inchH
2
O
7,50 mmHg
1 inchH
2
O
1,87 mmHg
0,25 kPa
2,53 cm H
2
O
Содержание Solar Blue
Страница 1: ......
Страница 9: ...8 2008 2020 by MMS B V...
Страница 19: ...18 2008 2020 by MMS B V Chapter 1...
Страница 47: ...46 2008 2020 by MMS B V Chapter 3...
Страница 75: ...74 2008 2020 by MMS B V Chapter 4...
Страница 81: ...80 2008 2020 by MMS B V Chapter 5...
Страница 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Страница 139: ...138 2008 2020 by MMS B V...
Страница 157: ...156 2008 2020 by MMS B V Appendix A...
Страница 165: ...164 2008 2020 by MMS B V Appendix C...
Страница 173: ...172 2008 2020 by MMS B V Appendix D...
Страница 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Страница 175: ...174 2008 2020 by MMS B V Appendix E...
Страница 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Страница 198: ......