Replacement of the batteries
LABORIE advises to replace the four rechargeable batteries with new ones at
least once every two years, or when the performance is getting low. Always
replace all batteries at the same time, with rechargeable NiMH AA batteries of
the same brand, type and capacity (
≥ 2300 mAh).
7.2.4
Connect the power adapter
The medical grade power adapter can be used (1) to charge the batteries in
the Bluetooth puller, and (2) to power the puller. During an investigation, it is
allowable to power the Bluetooth puller and to charge the batteries by using
the power adapter.
Best is to connect the power adapter when the Bluetooth puller is not used to
prevent batteries to become depleted. Completely depleted batteries will
shorten the life-span.
The procedure is as follows:
Connect the power adapter to the DC receptacle (see figure 7.2/[2]). Make
sure that the Bluetooth puller is connected to the correct power adapter
(PS-SBL). Do not use a power adapter other than which is specified. Doing
so may cause fire leading to serious accidents.
Press on the
Unlock slider
button on the control panel. Both the
Status
LED and the
Bluetooth
LED will flash once.
Charge the batteries for a period of at least 16 hours for a maximum
performance. Charging partially depleted batteries will require
approximately 8 to 10 hours of charge time. In parallel, you can register the
Bluetooth puller, test the puller and even perform patient investigations.
Batteries will get warm during charging. This will not damage the batteries and
is normal.
The DC receptacle on the Bluetooth puller should be kept clean for optimal
performance.
Содержание Solar Blue
Страница 1: ......
Страница 9: ...8 2008 2020 by MMS B V...
Страница 19: ...18 2008 2020 by MMS B V Chapter 1...
Страница 47: ...46 2008 2020 by MMS B V Chapter 3...
Страница 75: ...74 2008 2020 by MMS B V Chapter 4...
Страница 81: ...80 2008 2020 by MMS B V Chapter 5...
Страница 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Страница 139: ...138 2008 2020 by MMS B V...
Страница 157: ...156 2008 2020 by MMS B V Appendix A...
Страница 165: ...164 2008 2020 by MMS B V Appendix C...
Страница 173: ...172 2008 2020 by MMS B V Appendix D...
Страница 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Страница 175: ...174 2008 2020 by MMS B V Appendix E...
Страница 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Страница 198: ......