60
© 2008
– 2020 by MMS B.V.
Chapter 4
4.4.2
Pump Test
Choose the TAB
Pump
to test the pump of the Solar Blue module.
Figure 4.20
Pump test
The procedure is as follows:
Hang a bag (or bottle) of saline on the hook.
Install a pump tube in the pump.
Connect the 'input' side of the tube to the bag or bottle.
Connect the flowmeter via Bluetooth (when the hardware test Bluetooth is
used) or
Connect the flowmeter via a wired connection to an MMS-bus receptacle
(when the hardware test Solar (USB) is used). Make sure the power
adapter is connected to the Solar Blue module in this case.
Place the other side of the pump tube in the water container which is
positioned on the flowsensor. With this configuration, the 'inflow' can be
checked as the water is pumped from the bag. The flow measured by the
flowsensor can be compared to the selected pump speed.
Depending on the problem, select the desired pump speed.
Start the pump by clicking the
Start pump
button.
Содержание Solar Blue
Страница 1: ......
Страница 9: ...8 2008 2020 by MMS B V...
Страница 19: ...18 2008 2020 by MMS B V Chapter 1...
Страница 47: ...46 2008 2020 by MMS B V Chapter 3...
Страница 75: ...74 2008 2020 by MMS B V Chapter 4...
Страница 81: ...80 2008 2020 by MMS B V Chapter 5...
Страница 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Страница 139: ...138 2008 2020 by MMS B V...
Страница 157: ...156 2008 2020 by MMS B V Appendix A...
Страница 165: ...164 2008 2020 by MMS B V Appendix C...
Страница 173: ...172 2008 2020 by MMS B V Appendix D...
Страница 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Страница 175: ...174 2008 2020 by MMS B V Appendix E...
Страница 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Страница 198: ......