LABORIE Portascan3D Service Manual PTS3D-SM01 V3.0
35
STEP/ACTION
4
Test Scans
Test five scans and record average volume
readings
5
Click ‘SETUP’
Click ‘GAIN’
On
gain
interface, click
anywhere
in
green
zone
(right)
6
After
clicking
green
zone,
engineering
interface
will
appear
on
screen
7a.
Volume reading is
below
the specification:
Calibrate it by decreasing the base gain
i.e. System base gain = 18
129 ml phantom read 102 ml (7 ml below the specification)
Use “DW” bar adjust the base gain = 16
Click “SAVE”, “DEFAULT”, “BACK”, then “OK”
Scan again to verify volume reading reaches in the spec.