8
LABORIE Portascan3D Service Manual PTS3D-SM01
V3.0
D. Probe Scan shows only a vertical line
If you have another probe, use it for test.
1.
If system comes back to normal, this indicates that the
probe with no image had signal transmission
interruption. Please send the probe to the Service
Center for repair or cable replacement.
2.
If probe is changed and the system still has no image,
this indicates that all system failed. Please send your
unit and probe together to the Service Center for repair.
3.
If you do not have spare probe, please send unit and
probe together to the Service Center for repair.
E. Probe Connection Issues
Probe plug has a clip to lock the plug with the socket on the
unit. Insertion or removal needs to pull the unlock sleeve to right
to low the lock clip. If the plug is dirty, filled with dirt, gel, or oil, it
will be hard to connect or remove.
Fig. 4.3: Disassemble the plug and use 70% alcohol solution to
wash each component
4.3 DISPLAY
A. LCD Calibration
1.
Check if there are irrelevant materials at the surrounding
area of the screen and clear them off if there is.
2.
Calibrate the touch screen for trouble shooting. Touch
screen calibration guidance: Switch on the instrument
while pressing the probe button and operators will enter
Содержание PORTASCAN 3D