LABORIE Portascan3D Service Manual PTS3D-SM01 V3.0
31
Fig. 9.9: Two Board Unit
Fig. 9.10: Main Board Unit
A.
Main Board Troubleshooting
ISSUE
TWO BOARD
MAIN BOARD
Failure
of
scan
control
CPU fails. Replacing
digital board can solve
the problem.
Scan control IC (UD2, 74F08)
failed. Please contact the
service center to change the
chip.
Unit
does
not power
off
Contact service center
to replace: Q453 IC 9933
and U450 IC 2950
Contact service center to
replace: Q453 IC9933 and
U450 IC 2950